GE LOGIQ V5/LOGIQ V3
D
IRECTION 5496012-100, REVISION 3 BASIC SERVICE MANUAL
4 - 4 Section 4-3 - General Procedure
4-3-2-3 Check System Date and Time
A warning message “Please check the system date and time are correct” appears on the screen when
the system is powered on. This warning message appears for the possible reasons:
The system is not boot up for over 14 days.
The system time has been changed by 24 hours earlier than the current system time of last boot-up.
The BIOS time is changed by 24 hours earlier than the current system by resetting BIOS time,
replacing BIOS module or changing BIOS time.
This warning message is to remind the user to check the system date in case the system date and time
is incorrect.
Move the cursor to OK and press Cursor key on the control panel to select OK. The system enters
scanning mode.
Check the system date and time. If it is incorrect, follow below steps to reset the system date and time.
1.) Enter Utility -> System -> General -> Date/Time.
2.) Reset the system date and time.
3.) Select Apply and then select OK.
4.) Select Save.
Figure 4-3 Check system date and time message

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals