Site Preparations
3-16 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for the Ultrasound workstations
Contents:
• ‘Electrical Requirements - EchoPAC PC Turnkey
Workstations’ on page 3-16
• ‘Electrical Requirements - LOGIQworks Workstations’ on
page 3-16
Table 3-15: Electrical Requirements - EchoPAC PC Turnkey Workstations
Product Voltage Tolerance
Current or
Power
Consumption Frequency
All models
100 VAC
±10%
8 A
50-60 Hz
110 VAC 8 A
230 VAC 4 A
Table 3-16: Electrical Requirements - LOGIQworks Workstations
Product Voltage Tolerance
Current or
Power
Consumption Frequency
All models 110 - 230 VAC ±10% 310 to 543 W
(depending on
workstation
model)
50-60 Hz