System Setup
3-48 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
To add a server in the Server Config list
Follow this procedure to add a server to the list:
1. Select Add.
Figure 3-16. Server Configuration
2. Add the server's name in the Server Name field.
3. Add the server’s IP address in the IP-Address field.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right of the
Check button indicates the result:
• A green check mark next to the Check button indicates
that the IP-Address was found on the network.
• A red mark indicates that the IP address was not found.
Troubleshooting tips:
• No network connection between the Ultrasound system
and the server.
• The wrong IP Address.
To modify the setup for a server in the list
NOTE: Both the pre-defined servers and new servers can be modified.
Follow this procedure to modify the setup for a server in the
list:
1. Highlight the server you want to modify.
2. Select Modify.
3. Update the information as needed.
4. Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right for the
Check button indicates the result:
• A green check mark next to the Check button indicates
that the IP-Address was found on the network.
• A red mark indicates that the IP address was not found.
Troubleshooting tips:

Table of Contents

Other manuals for GE Vivid E80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E80 and is the answer not in the manual?

GE Vivid E80 Specifications

General IconGeneral
BrandGE
ModelVivid E80
CategoryMedical Equipment
LanguageEnglish

Related product manuals