Care and Maintenance
10-4 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Maintenance task schedule
How often should maintenance tasks be performed?
The Care and Maintenance task schedule specifies how often
your Vivid E80/E90/E95 should be serviced and outlines items
requiring special attention.
NOTE: It is the customer’s responsibility to ensure the Vivid E80/E90/
E95 care and maintenance is performed as scheduled in order
to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowledge of
your Vivid E80/E90/E95 and can best provide competent,
efficient service. Contact GE for coverage information and/or
price for service.
The service procedures and recommended intervals shown in
the Care and Maintenance Task Schedule assumes that you
use your Vivid E80/E90/E95 for an average patient load (10-12
per day) and not use it as a primary mobile Ultrasound system
which is transported between diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load,
then it is recommended to increase the periodic maintenance
frequencies.
Table 10-1: Customer Care Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Per Facilities
QA Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders •
Clean Air Filters • more frequently depending on
your environment
Inspect AC Mains Cable • Mobile Ultrasound system:
Check Weekly