LCD Monitor and LCD Arm parts replacement
Vivid E80/E90/E95 – Service Manual 8-29
GC091052 Rev. 3
LCD Monitor and LCD Arm parts
replacement
Replacing the 19” LCD Monitor assembly
Read and follow
Preparations
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Move the Top Console to its lower position.
3. Disconnect the Mains Power Cable from the wall outlet.
4. Disconnect all probes and I/O cabling.
Related information
• ‘Power shut down’ on page 4-5
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.