Replacement Procedures
8-20 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Plate connectors
w
/guide removal
The plate connectors
w
/guide is attached with hatches.
Perform these steps to remove the plate connectors
w
/guide:
1. Use a flat bladed screwdriver to loosen the hatches.
2. Remove the Plate Connectors w/Guide
Plate connectors
w
/guide installation
Perform these steps to install the plate connectors
w
/guide:
1. Position the plate and snap it into position.
2. Install front cover.
3. Install Foot rest bumper.
4. Install top cover.
5. Install side covers.
Verification
Perform the following steps to verify that the product is
functioning as intended after this replacement:
1. Connect cables and probes you removed earlier
2. Power up the Ultrasound system to verify that it operates as
intended.