Replacement Procedures
8-64 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
3. Disconnect all probes and external I/O cabling.
4. Remove the LCD Rear Cover.
5. Remove the LCD Monitor assembly.
6. Remove the LCD Arm assembly.
7. Remove the Rotary buttons below the Touch screen and the
Volume button.
8. Remove the Operating Panel, Upper.
9. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
10. Remove the Bulkhead Board.
11. Remove the Bulkhead Plate.
12. Remove the Upper Frame Assembly.
13. Remove the Lower Frame Assembly.
Remove the XY Mechanism
Follow this procedure to remove the XY Mechanism:
1. From the rear side of the Ultrasound system, remove the
four (4x) fixing screws.
2. Remove the XY mechanism.
Install the XY Mechanism
Follow this procedure to install the XY Mechanism:
1. Position the XY mechanism so it aligns with the holes for the
four fixing screws.
2. Install the four fixing screws.
3. Install the Lower Frame Assembly.
NOTE: Ensure that Bumper Z enters the slide at the back, before
tighten the screws.
4. Install the Upper Frame Assembly.
5. Install the Bulkhead Plate.
6. Install the Bulkhead Board.
7. Install the Operating Panel, Lower.
8. Install the Operating Panel, Upper.
9. Install the LCD Arm assembly.
10. Install the LCD Monitor assembly.
11. Install the LCD Rear Cover.

Table of Contents

Other manuals for GE Vivid E80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid E80 and is the answer not in the manual?

GE Vivid E80 Specifications

General IconGeneral
BrandGE
ModelVivid E80
CategoryMedical Equipment
LanguageEnglish

Related product manuals