2.2.2 Remove the expiratory cassette
SERVO-i
SERVO-i
Lift the locking handle and pull out the
patient unit.
Press the button on the expiratory cassette,
tilt it upwards and remove.
SERVO-s
Press the locking arm on the rear to tilt the
User Interface.
Press the button on the expiratory cassette,
tilt it upwards and remove.
WARNING! After removing the expiratory
cassette, do not pour any fluid into the
expiratory cassette compartment. Avoid
contact with electrical connectors.
Important: The expiratory cassette can be
exchanged between different SERVO-i/s
ventilator systems. The ventilator may be
used immediately* by connecting a clean
expiratory cassette.
* After replacing the expiratory cassette, a
Pre-use check must be performed (except in
emergencies, when it should be performed as
soon as possible).
10
SERVO-i/s, User's Manual