2.3 Wiping and discarding
D2D1
Dismantle
Wipe off/discard
Rinse
Disinfectant
Rinse
Washer-
Disinfector
Dry
Assemble
• Wipe the ventilator and all removable parts
with a soft lint-free cloth moistened in soap
& water or detergent-based disinfectant.
3M Respirator Cleaning Wipes-
- Clorox Bleach Germicidal Wipes (may
cause discoloration)
- Super Sani-Cloth Germicidal Wipes (may
cause discoloration)
- Virkon S
- Accel TB (Virox 5)
Note: In case of more contaminated
surfaces, use ethyl alcohol (70%) or
isopropyl alcohol (70%). Avoid contact with
electrical contacts.
CAUTION: Do not immerse battery, CO
2
analyzer, Edi module or Y sensor module
(SERVO-i only) in any fluid.
• Discard disposable items:
- Servo Duo Guard viral/bacterial filter
- Servo Humidifier/HME
- Y sensors
- disposable patient tubing
- disposable airway adapter
- Edi catheters
- Aerogen Solo nebulizer
- disposable CO
2
adapter
Notes:
• Single use (i.e. disposable) items shall
not be reused. Cleaning of single use
items can damage the item and
compromise its intended use. Reuse of
single use items increases the risk of
spreading infections and compromises
the durability of the item.
Regularly check that the fan filter looks clean
(i.e. black). If dusty, remove (snap off/snap on)
and rinse in water. Shake out, ensuring that
the filter is free from excess water.
This is all the cleaning that is required when a
Servo Duo Guard filter (or equivalent) is used.
SERVO-i/s, User's Manual
11
| Routine cleaning |
2 |