3.1 Preventive maintenance
Maquet recommends that the equipment is
inspected regularly.
A preventive maintenance, according to
corresponding chapter in the Service manual,
must be performed by authorized personnel
at least once every year as long as the unit is
not used more than normal. Normal operation
during one year is estimated to correspond to
approximately 5000 hours of operation. The
current operating time and time to next
preventive maintenance is presented under
the Status menu on the User Interface.
Important: It is recommended that a regular
cleaning and an extended cleaning of the
inspiratory channel should be performed
before carrying out preventive maintenance.
The extended cleaning must be done by
trained personnel only.
3.2 O
2
sensor
No regular maintenance required.
3.3 O
2
cell
Regularly check the cell status in the Status
menu. The O
2
cell should be exchanged if
<10% estimated remaining capacity is
indicated in the Status menu. Refer to
SERVO-i/s User's manuals for O
2
cell
adjustment.
WARNING! The sealed unit of the O
2
cell
contains a caustic liquid which may cause
severe burns to the skin and eyes. In case
of contact, immediately wash the affected
area continuously with water for at least 15
minutes and seek medical attention,
especially if the eyes are affected.
Important:
Make sure the O
2
cell is for the
SERVO-i/s models.
Replacement of the O
2
cell and filter and
extended cleaning may only be
performed by trained personnel.
SERVO-i/s, User's Manual
27
| Maintenance |
3 |

Other manuals for Getinge Maquet SERVO-i

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-i and is the answer not in the manual?

Getinge Maquet SERVO-i Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-i
CategoryMedical Equipment
LanguageEnglish

Related product manuals