2x
• Connect the O
2
sensor (two "clicks" are
heard).
• Connect the O
2
sensor connector.
3.4.6 Assembling
• Re-assemble according to instructions
given on page 30.
• Note on a log sheet that an extended
cleaning of the inspiratory channel has been
performed.
• Perform a Pre-use check (see SERVO-i/s
User's Manuals).
Important: The accuracy of the
measurement is dependent of the quality
(oxygen content) of supplied gases during
Pre-Use Check.
3.5 Accessories
Fischer & Paykel Humidifier MR850
(optional)
Refer to the operating manual for the Fisher
& Paykel Humidifier MR850.
Disposable parts
Use disposable and spare parts from Maquet
only. All disposable parts must be discarded
according to hospital routine and in an
environmentally safe way.
34
SERVO-i/s, User's Manual
| 3
| Maintenance |