8 Opening and closing the lid
8.1 Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on “Emergency release”.
Press the key and open the lid. The symbol " " (lid open) illuminates in the rotation indicator .
8.2 Closing the lid
Do not bang the lid shut.
Place the lid and press the lid grip down slightly. The symbol "
" (lid closed) illuminates in the rotation indicator
.
9 Installation and removal of the rotor
A
B
C
D
Clean the motor shaft (C) and the rotor drilling (A), and lightly grease the motor shaft
afterwards. Dirt particles between the motor shaft and the rotor hinder a perfect seating
of the rotor and cause an irregular operation.
Place the rotor vertically on the motor shaft. The driver areas (D) of the motor shaft
must be located in the groove (B) of the rotor. The alignment of the groove is labelled
on the rotor.
Tighten the rotor tension nut with the supplied wrench by turning in a clockwise
direction.
Check the rotor for firm seating.
Loosening the rotor: Loosen the tensioning nut by turning anti-clockwise. Turn the
tensioning nut until the rotor is able to be lifted from the motor shaft.
10 Loading the rotor
Standard centrifuge containers of glass will not stand RCF values exceeding 4000 (DIN 58970, pg. 2).
Check the rotor for firm seating.
The rotors must be loaded symmetrically. For authorised combinations see Chapter "Anhang/Appendix, Rotoren
und Zubehör/Rotors and accessories".
Always fill the centrifuge containers outside of the centrifuge.
The maximum filling quantity for the centrifuge containers specified by the manufacturer must not be exceeded.
In order to maintain the weight differences within the centrifuge container as marginal as possible, a consistent
fill level in the containers is to be heeded.
The weight of the permissible filling quantity is specified on each rotor. This weight may not be exceeded.
11 Closing biosafety systems
In order to guarantee that it is sealed, the lid to the biosafety system has to be firmly closed.
In order to prevent the packing ring from twisting when opening and closing the cover, the packing ring must
be lightly rubbed with talcum powder or a rubber care product.
For further details of available biosafety systems see chapter "Anhang/Appendix, Rotoren und Zubehör/Rotors
and accessories". If in doubt, you should obtain relevant information from the manufacturer.
Lid with screw top, without hole in turning handle:
Close the lid firmly by hand by turning in clockwise direction.

Table of Contents

Other manuals for Hettich MIKRO 120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich MIKRO 120 and is the answer not in the manual?

Hettich MIKRO 120 Specifications

General IconGeneral
BrandHettich
ModelMIKRO 120
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals