9 Opening and closing the lid 打开与关闭盖子
9.1 Opening the lid 打开盖子
只有当离心机的开关打亮起并且转子是静止时,盖子才可打开。如不能打开,请看“盖锁紧急
开启” 这一章节。
Press the button
The lid opens automatically and the left LED in the button
go off.
按钮。盖子会自动打开,按钮 左边的 LED 灯会熄灭。
9.2 Closing the lid 关闭盖子
不要把手指放在盖子及离心室之间。
不要“砰”地一声把盖子关上。
Place the lid and lightly press down the front edge of the lid. The locking action is effected by motor.
The left LED in the button
lights up.
放好盖子,轻下压盖子前面边缘,盖子会被自动锁上。按钮
左边的 LED 灯会亮起。
10 Installation and removal of the rotor 转子的安装与卸载
Clean the motor shaft (C) and the rotor drilling (A), and lightly grease the motor
shaft afterwards. Dirt particles between the motor shaft and the rotor hinder a
perfect seating of the rotor and cause an irregular operation.
清洁发动机轴(C)及装转子的孔(A)并随后给发动机轴涂上一层薄润滑油。
发动机轴及转子之间的脏物会影响转子的放置,造成运行失调。
Place the rotor vertically on the motor shaft. The motor shaft dog (D) has to fit in the rotor slot (B).
The alignment of the groove is labelled on the rotor.
把转子垂直地装在发动机轴上。 发动机轴的驱动区域(D)必须要位于转子的凹槽里面(B。在
转子上标出了与凹槽相对应的位置。

Other manuals for Hettich MIKRO 220

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich MIKRO 220 and is the answer not in the manual?

Hettich MIKRO 220 Specifications

General IconGeneral
BrandHettich
ModelMIKRO 220
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals