EasyManuals Logo
Home>Honeywell>Measuring Instruments>954 SmartServo FlexLine

Honeywell 954 SmartServo FlexLine User Manual

Honeywell 954 SmartServo FlexLine
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Part No: 4417341 Revision 10 Installation Manual 36
954 SmartServo FlexLine
HART Transmitters
Stand-alone HART SmartView Display
Refer to HART Communication Foundation document number HCF_SPEC-54 (FSK
Physical Layer Specification) for detailed length calculations.
Connection Table
The I/O shall not be accessible or shall be isolated in the end use. Ensure the end user cannot touch these
wires when energized.
Cable specifications
Type
Twisted pair, shielded. R
max
= 25 / line
Maximum length
1000 meter
Label
Parameters
24. V_Loop
[Ex ia] IIB The I/O shall not be accessible or shall be isolated in the end use.
Uo = 23.1 V Ensure the end user cannot touch these wires when energized.
Io = 124 mA
Po = 0.6 W
Co = 1.0 μF
Lo = 9 mH
25. GND_Loop
Cable specifications
Type
Twisted pair, shielded. R
max
= 25 / line
Maximum length
1000 meter
Label
954 SmartServo
FlexLine
Parameters
954 SmartServo
FlexLine
Label
HART SmartView
Parameters
HART SmartView
24. V_loop
[ia Ga] IIB T6 Ga/ Gb
2. Loop +
[Ex ia] IIB T4 Gb
25. GND_Loop
Uo = 23.1 V
1. Loop -
Ui = 30 V
Io = 124 mA
Ii = 200 mA
Po = 0.6 W
Pi = 0.7 W
Co = 1.0 μF
Ci = 10 nF
Lo = 9 mH
Li = 16 μH

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell 954 SmartServo FlexLine and is the answer not in the manual?

Honeywell 954 SmartServo FlexLine Specifications

General IconGeneral
BrandHoneywell
Model954 SmartServo FlexLine
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals