13
• LOW PRESSURE (Refer to “LOW PRESSURE” on
page 12).
• HIGH PRESSURE (Refer to “HIGH PRESSURE” on
page 12).
•PUMP FAULT (Refer to “PUMP FAULT” on
page 13).
•
POWER (Refer to “POWER Fail” on page 13).
For all alarm conditions except Power Fail, once the alarm condition has
been detected and displayed, it can only be cancelled by switching the pump
unit off and then back on.
Refer to Section 8, Page 26 “Troubleshooting” for
possible causes of the above alarm conditions.
PUMP FAULT The PUMP FAULT indicator is illuminated when an
internal pump malfunction is detected.
The fault can only be rectified by carrying out a service
on the pump.
POWER Fail The POWER indicator will flash when a mains power
failure has been detected.
The alarm will continue until the mains power is
resumed or the pump is switched off using the
POWER
switch on the pump control panel.
Service Indicator The “telephone” symbol will be illuminated after a
set number of running hours to indicate that the pump is
ready for a service.
This service period is set to 12 months.
The pump will continue to operate normally even when the symbol is
illuminated.
Mattress Controls All
Nimbus 3 and Nimbus 3 Professional mattresses
have the following two controls, situated at the foot end
of the mattress:
Transport ControlCPR Control
NORMAL
TRANSPORT