19
CPR Control
In the event of a patient suffering cardiac arrest and CPR
needing to be administered:
To Activate the CPR 1. Lift the red CPR handle at the foot end of the
mattress.
2. Turn the handle counterclockwise.
3. Pull the handle away from panel.
4. The grey triangular seal will rotate and the air will
exhaust from the mattress. The torso area of the
patient will bottom out in less than 10 seconds.
To Reset the CPR 1. Turn the grey triangular seal clockwise and push
onto the connectors.
2. Turn the red handle clockwise.
3. Fold the handle flat to lock in position.
IMPORTANT
IN THE EVENT OF CARDIAC ARREST.