EasyManuals Logo
Home>iM3>Media Converter>42-12

iM3 42-12 User Manual

iM3 42-12
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #6 background imageLoading...
Page #6 background image
6
F. OPERATING TECHNIQUE
It is important to use the 42-12 correctly to benefit from its unique features.
1. Always use the tip parallel to the tooth, with the side of the tip on the tooth.
2. Do not use the tip as a pick or chisel at right angles to the tooth, as this will
scratch the enamel. Animal enamel is considerably thinner than human
tooth enamel, and therefore more prone to damage.
3. Apply lightly to the tooth surface, sweeping back and forth as though one
was painting the teeth with a paintbrush. Keep the tip moving. Use the tip
to probe deep pockets, but always parallel to the tooth surface.
4. As the tip operates in a 360º circular motion, easy access is guaranteed,
as all surfaces are active and there are no dead zones.
5. The circular motion actually brushes the teeth, rather than hammering
them, and produces more of a polishing effect, rather than a scratching
effect, as with other units.
NOTE: Increased pressure will not improve efficiency or the time taken to
complete the procedure.
6. Coolant flow control (11). Adjust the spray to achieve a fine mist. The use
of CLS is recommended in animals. Correct dilution is important (Read
Manufacturer‟s Recommendations) as over dilution may cause the
chlorhexadine to deposit small crystals.
NOTE: The use of any other solution other than CLS or distilled water in the
42-12 will void all warranties. CLS has been used successfully in air driven
Dental Units and ultrasonic scalers since 1993.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the iM3 42-12 and is the answer not in the manual?

iM3 42-12 Specifications

General IconGeneral
BrandiM3
Model42-12
CategoryMedia Converter
LanguageEnglish

Related product manuals