tina83e1-d (2018-12) 11
Output impedance
2 × 4.7 , short-circuit proof
Load impedance
10 k
, short-circuit proof
Step response time
>1×10
-6
mbar
1×10
-6
1×10
-8
mbar
pressure dependent
<100 ms
1 s
Gauge identification
3MAx-xxx-000N
1)
3MAx-xxx-000Q
3MBx-xxx-000P
100 k referenced to supply
common
85 k referenced to supply
common
Status signal (digital output)
FCC receptacle
Current rating
100 mA
High voltage is ON +14.5+30 V (dc)
(depending on supply voltage)
High voltage is OFF
0 V (dc)
D-sub receptacle
Supply voltage
30 V (dc)
Current rating
100 mA (sink)
High voltage is ON
0 V (dc)
High voltage is OFF open
High voltage cut-in, low active (digital input)
High voltage ON
High voltage OFF
<2.5 V (dc)
>4.0 V (dc)
High voltage cut-in, high active (digital input)
High voltage ON
High voltage OFF
>11.0 V (dc)
< 5.0 V (dc)
1)
Not suited for operation with a VGC40x / VGC50x controller.