EasyManuals Logo
Home>Inficon>Measuring Instruments>Gemini MPG504

Inficon Gemini MPG504 User Manual

Inficon Gemini MPG504
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
12 tina83e1-d (2018-12)
Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments, or control devices that conform
to the requirements of a grounded protective extra-
low voltage (PELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.
2)
Supply voltage
at the gauge
3)
+14.5+30 V (dc)
Ripple
1 V
pp
Power consumption MxG50x
≤2 W
Fuse to be connected
2)
1 AT
High voltage in the measuring chamber
Ignition voltage
Operating voltage
≤4.5 kV
≤3.3 kV
3Mxx-x0x-xxxx
3Mxx-x1x-xxxx
high current
low current
Electrical connection
3Mxx-xxx-00xx
3Mxx-xxx-0
1
xx
1)
FCC 68, 8-pin
D-sub, 9-pin
Sensor cable
FCC connector
D-sub connector
8-pin, shielded
9-pin, shielded
Cable length (FCC only) ≤50 m
(0.14 mm
2
/conductor)
2)
INFICON controllers fulfill this requirement.
3)
The minimum voltage of the power supply unit must be increased
proportionally to the length of the sensor cable.