Page #119 background imageLoading...
Page #119 background image
TraceRelated Command(IT6500C)
Copyright© Itech Electronics Co., Ltd. 108
Query syntax
TRACe:DELay? [MINimum|MAXimum|DEFault]
Returns
<NR3>
TRACe:TIMer
This command is used to select cache time interval.
Command syntax
TRACe:TIMer < NRf>
Arguments
0.00002 to 3600s|MINimum|MAXimum|DEFault
Unit
s
Example
TRAC:TIM 0.1
Query syntax
TRACe:TIMer? [MINimum|MAXimum|DEFault]
Returns
<NR3>
TRACe:POINts:ACTual?
This command is used to select number of actual readings in the cache.
Command syntax
TRACe:POINts:ACTual?
Returns
<NR1>
TRACe:CLEar:AUTO[:STATe]
This command is used to select whether to automatically clear cache data.
When Auto Clear is enabled, when saving starts, buffer will be automatically
cleared; when it is disabled, the new reading will be saved to the buffer area
following the previous data till the buffer is full or when saving stops.
Command syntax
TRACe:CLEar:AUTO[:STATe] <bool>

Table of Contents