Load Commands
Copyright © Itech Electronic Co., Ltd. 34
Command syntax:
WAVE:TRIGger:SLOPe <POSitive|NEGative|ANY>
Parameter:
POSitive|NEGative|ANY
Query Syntax:
WAVE:TRIGger:SLOPe?
Returned value:
POSitive|NEGative|ANY
WAVE:TRIGger:MODE
This command is used to set the trigger mode when take waveform as trigger
source.
Command syntax:
WAVE:TRIGger:MODE <AUTO|NORMal>
Parameter:
AUTO|NORMal
Query Syntax:
WAVE:TRIGger:MODE?
Returned value:
AUTO|NORMal
WAVE:TRIGger:DELay:TIME
This time is used to set the trigger delay time when take a waveform as trigger
source. (Note: the delay time setting should be less than the time width of the
whole screen.
Command syntax:
WAVE:TRIGger:DELay:TIME <NRf>
Parameter:
<NRf>
Unit:
s