Calibration Commands
Copyright © Itech Electronic Co., Ltd. 53
Chapter6 Calibration Commands
CALibrate:SECure
This command is used to enable or disable calibration mode.
The calibration mode must be enabled before the load will accept any other
calibration commands. The first parameter specifies the enabled or disabled
state. The second parameter is the password. It is required if the calibration
mode is being enabled and the existing password is not 0. If the password is
not entered or is incorrect, an error is generated and the calibration mode
remains disabled. The query statement returns only the state, not the password.
Whenever the calibration state is changed from enabled to disabled, any new
calibration constants are lost unless they have been stored with
CALibrate:SAVE.
Command syntax:
CALibrate:SECure[:STATe] <bool> [,<SRD>]
Parameter:
0 | 1 | OFF | ON [,<password>]
For example:CAL:SEC 0, N3301A CAL:SEC ON
Reset value:
ON
Query Syntax:
CALibrate:SECure[:STATe]?
Returned value:
<NR1>
Relevant Command:
CAL:SAV CAL:INIT
CALibrate:INITital
This command can only be used in calibration mode. It restores factory
calibration constants from nonvolatile memory.
Command syntax:
CALibrate:INITial
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com

Table of Contents

Other manuals for ITech IT8625

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8625 and is the answer not in the manual?

ITech IT8625 Specifications

General IconGeneral
BrandITech
ModelIT8625
CategoryTest Equipment
LanguageEnglish

Related product manuals