Operations introduction
Copyright © Itech Electronic Co., Ltd. 51
Wdc=0.00W
3. Press + to enter configuration menu
Sync On Set
Von
Measure Rate
4. Press to select <Von>, press to enter. First set <Von point>,
press key to enter, input the programmed value, such as 1V, press
key to confirm
Von point
Von Latch
Exit
5. Press to select <Von Latch>, press key to enter, select
Latch to enable on/off state, such as On, press key to confirm
6. Press key to exit
7. Open the load’s input
When in remote control mode USB / RS232 / Ether-net), you can send SCPI
command VOLT:ON <n> to set Von value; send VOLT:LATch ON to enable Von
LATCH function.
4.11 Protection functions
Load has the following protection functions: over voltage protectionOVP,over
current protectionOCP, over power protection(OPP), over temperature
protection(OTP), reverse voltage alarm (LRV/RRV.
The mainframe will act appropriately once any of the above protection is active.
You can press any button on front panel to restore the protection function. For
example, the electronic load come into over temperature protection, the buzzer
will alarm, the input will automatically shut down and Main frame VFD will
display OTP.
4.11.1 Over voltage protection (OVP)
If the OVP circuit has triggered, buzzer alarm, the main frame screen will
displays (OVP), the condition will remain until they are reset.
Operations to clear the OVP state:
Check whether the input voltage is in the load’s rated voltage or the
programmed protecting voltage ranges. If it is outside of the range, please
disconnect the instrument under test. Press any key on the front panel (or send
commandPROTection:CLEar), the (OVP) displayed on the front panel will
disappear, load exits OVP protection state.
4.11.2 Over current protection (OCP)
The electronic load includes both hardware and software over current
protection features.
Hardware OVP: load’s maximum input current will be limited at about 110%
of the current range, once the hardware OCP is triggered, the status
registers OC bit will be set; when the hardware OCP is removed, the
status register’s OC bit will be reset. Hardware over current protection
won’t affect load’s on/off state.
Software OCP: users can set load’s software OCP value, steps: +
ShiftShift
55
EnterEnter
EnterEnter
EnterEnter
EnterEnter
EnterEnter
EscEsc
ShiftShift

Table of Contents

Other manuals for ITech IT8731

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8731 and is the answer not in the manual?

ITech IT8731 Specifications

General IconGeneral
BrandITech
ModelIT8731
CategoryTest Equipment
LanguageEnglish

Related product manuals