EasyManuals Logo

Keithley 2002 Getting Started Manual

Keithley 2002
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background image
SCPI signal oriented commands
Signal oriented commands are used to acquire readings using a set of high-level instructions
to control the measurement process. The two most commonly used commands are
:FETCh?
and
:MEASure?
. Details on all signal oriented commands are covered in Section 3 in the User’s Man-
ual.
:FETCh?
This query command is used to request the latest post-processed reading. After sending this
command and addressing the Model 2002 to talk, the reading will be sent to the computer. This
command does not trigger a measurement, it simply requests the last available reading.
:MEASure[:<function>]?
In general, this command places the instrument in the “one-shot” measurement mode (ARM
indicator off), triggers a single measurement, and then performs a
:FETCh?
.
If you include the measurement function with this command, the instrument will first go to
the specified function and then perform the one-shot measurement and query, as follows:
:MEASure?
Currently selected function
:MEASure:VOLTage:DC?
DCV function
:MEASure:VOLTage:AC?
ACV function
:MEASure:CURRent:DC?
DCI function
:MEASure:CURRent:AC?
ACI function
:MEASure:RESistance?
2 function
:MEASure:FRESistance?
4 function
:MEASure:FREQuency?
FREQ function
:MEASure:TEMPerature?
TEMP function
IEEE-488 Bus Operation 3-11

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A