EasyManuals Logo
Home>Keithley>Multimeter>2002

Keithley 2002 Calibration Manual

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
C
Calibration Messages
C-1
C.1 Introduction
This appendix lists calibration errors that may occur during
calibration as well as the :CAL:PROT:DATA? query
response.
C.2 Error summary
Table C-1 summarizes Model 2002 calibration errors. These
errors are displayed on the front panel and may be read over
the bus by using the :SYST:ERR? query. The query response
is the error number and the error message enclosed in quotes.
For example,
+444,"Cal step generated invalid data"
Error generation
After each calibration command is sent, the Model 2002 will
perform the appropriate calibration step and calculate perti-
nent calibration constants. If an error is detected, the instru-
ment will generate the appropriate error message, as
summarized in Table C-1.
Temperature drift error
An "Excessive temp drift during cal" error (+519) may occur
if the internal temperature of the Model 2002 drifts exces-
sively during calibration. The Model 2002 measures its inter-
nal temperature when the ":CAL:PROT:INIT" command is
sent and then again when the ":CAL:PROT:SAVE" com-
mand is sent at the end of the calibration procedure. If the in-
ternal temperature drift is excessive, the ERR annunciator
will turn on, and the "Excessive temp drift during cal" mes-
sage will be placed in the error queue.
Note that this error condition does not prevent calibration
constants derived during calibration from being saved; rath-
er, it is intended to flag excessive temperature drift, possibly
caused by insufficient warm-up.
NOTE
Placement of the OPTION SLOT cover af-
fects the internal temperature of the Model
2002. To achieve T
CAL
±1°C specifica-
tions, the OPTION SLOT cover must be in
the same position (on or off) as when the
Model 2002 is to be used.
Invalid calibration data error
A "Cal step generated invalid data" error (+444) will be gen-
erated when the ":CAL:PROT:SAVE" command is executed
if a previous calibration step failed. This error will be
repeated for each ":CAL:PROT:SAVE" command until the
calibration step in question is successfully completed.
C.3 Calibration data query response
Table C-2 lists the response to the :CAL:PROT:DATA?
query. The response is a string of ASCII floating-point num-
bers separated by commas, and it is not affected by the FOR-
MAT command. Constants listed in Table C-2 are shown in
the order they are sent.

Table of Contents