EasyManuals Logo

Keithley 2002 Calibration Manual

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background image
Calibration
2-11
(Send the actual calibration value in the range of 9.5V to
20.5V if you are using a different voltage.)
7. Wait until the Model 2002 finishes this step before going
on.
Step 4: Ohms calibration
1. Set the calibrator output to 1M, and turn external sense
on.
NOTE
External sensing (4-wire ohms) should be
used when calibrating all resistance
ranges.
2. Send the following command to the Model 2002:
:CAL:PROT:DC:OHM1M <value>
Here, <value> is the actual calibrator resistance value. For
example, if the calibrator resistance is 1.002M, the com-
mand would appear as follows:
:CAL:PROT:DC:OHM1M 1.002E6
(The allowable range for this parameter is from 475E3 to
1.025E6.)
3. Wait until the Model 2002 finishes 1M calibration be-
fore continuing.
4. Set the calibrator resistance to 100k, and make sure
external sense is still turned on.
5. Repeat steps 2 and 3 for each of the remaining ohms cal-
ibration points as shown in Table 2-5.
Step 5: Amps calibration
1. Connect the calibrator to the AMPS and INPUT LO
jacks, as shown in Figure 2-3.
2. Set the calibrator output to 200.000µA, and place the
unit in operate.
3. Send the following command to the Model 2002:
:CAL:PROT:DC:A200U 200E-6
If you are using a different calibration value, be sure to sub-
stitute that value for the parameter shown above. (The allow-
able range is from 95µA to 205µA.)
Wait for the instrument to complete this step before
continuing.
4. Repeat steps 2 and 3 for the remaining amps calibration
points shown in Table 2-6.
Step 6. Open-circuit calibration
1. Disconnect all test leads from the Model 2002 INPUT
and AMPS jacks.
2. Send the following command to the instrument:
:CAL:PROT:DC:OPEN
3. Wait until open-circuit calibration is complete before
going on to the next step.
Table 2-5
Ohms calibration summary
Cal
point
Nominal
resistance Allowable range Command*
2M
200k
20k
2k
200
20
1M
100k
19k
1.9k
190
19
475k to 1.025M
95k to 205k
9.5k to 20.5k
0.95k to 2.05k
95 to 205
9.5 to 20.5
:CAL:PROT:DC:OHM1M 1E6
:CAL:PROT:DC:OHM200K 100E3
:CAL:PROT:DC:OHM20K 19E3
:CAL:PROT:DC:OHM2K 1.9E3
:CAL:PROT:DC:OHM200 190
:CAL:PROT:DC:OHM20 19
* Nominal resistance values shown. Use exact calibrator resistance value for command parameter.

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
Resolution8.5 digits
DC Voltage Accuracy±0.0006%
Maximum DC Voltage1000 V
Maximum AC Voltage750 V
Maximum DC Current3 A
Maximum AC Current3 A
ConnectivityGPIB, RS-232
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
Model2002
CategoryMultimeter
InterfaceGPIB, RS-232
AC Current Accuracy±0.1%
Frequency Range0.1Hz to 300kHz
DisplayVFD
TypeDigital
Power Supply100-240V AC, 50/60Hz
Current Range (DC)100 μA to 3 A
Current Accuracy (DC)±0.01%
Current Range (AC)100 μA to 3 A
Current Accuracy (AC)±0.1%
DC Current Range100 μA to 3 A
AC Current Range100 μA to 3 A

Related product manuals