EasyManuals Logo
Home>Keithley>Multimeter>2002

Keithley 2002 Calibration Manual

Keithley 2002
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #46 background imageLoading...
Page #46 background image
Calibration
2-17
INPUT = 20.0000000 V
(At this point, you can use the cursor and range keys to set
the calibration voltage to a value from 9.5 to 20.5V if your
calibrator cannot source 20V.)
9. Press ENTER. The instrument will display the
following message to indicate it is performing 20V DC
calibration:
Performing 20 VDC Calibration
Step 5: Ohms calibration
1. After completing 20VDC calibration, the instrument
will display the following:
CONNECT 1 M 4W
2. Set the calibrator output to 1.00000M, and make sure
that external sense is turned on.
NOTE
Be sure that the calibrator external sense
mode is turned on when calibrating all re-
sistance ranges.
3. Press ENTER, and note that the Model 2002 displays
the resistance calibration value:
INPUT = 1.0000000 M
4. Using the cursor and range keys, set the resistance value
displayed by the Model 2002 to the exact resistance val-
ue displayed by the calibrator. (The allowable range is
from 475k to 1.025M.)
5. Press ENTER, and note that the instrument displays the
following during 1M calibration:
Performing 1 M Calibration
6. After completing 1M calibration, the instrument will
display the following:
CONNECT 100 k 4W
7. Set the calibrator output to 100k, and make sure that
external sense is turned on.
8. Press ENTER, and note that the Model 2002 displays
the resistance calibration value:
INPUT = 100.00000 k
9. Using the cursor and range keys, set the resistance value
displayed by the Model 2002 to the exact resistance val-
ue displayed by the calibrator. (The allowable range for
this parameter is from 95k to 205k.)
10. Press ENTER to complete the 200k calibration step.
11. Repeat steps 7 through 10 for the 20k, 2k, 200, and
20 ranges in that order. Be sure the set the calibrator
and Model 2002 to the correct resistance value as
follows:
Step 6: DC amps calibration
1. After ohms calibration is completed, the instrument will
prompt you for the first DC amps calibration step:
CONNECT 200 µADC
2. Connect the DC amps calibrator to the AMPS and
INPUT LO terminals (see Figure 2-3).
3. Set the calibrator output to 200.000µA, and make sure
the unit is in operate. (The allowable range is from 95µA
to 205µA.)
4. Be sure that the displayed current matches the calibra-
tion value, then press ENTER to complete this calibra-
tion step.
5. Repeat steps 3 and 4 for the remaining amps calibration
points as follows:
Step 7: Open-circuit calibration
1. At this point, the instrument will display the following
message advising you to disconnect test leads:
OPEN CIRCUIT INPUTS
Calibration
step
Calibration
value* Allowable range
2M
200k
20k
2k
200
20
1M
100k
19k
1.9k
190
19
475k to 1.025M
95k to 205k
9.5k to 20.5k
0.95k to 2.05k
95 to 205
9.5 to 20.5
* Nominal values shown. Use exact calibrator value.
Calibration
step
Calibrator
current Allowed range
200µA
2mA
20mA
200mA
2A
200.000µA
2.00000mA
20.0000mA
200.0000mA
1.00000A
95µA to 205µA
0.95mA to 2.05mA
9.5mA to 20.5mA
95mA to 205mA
0.95A to 2.05A

Table of Contents

Other manuals for Keithley 2002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2002 and is the answer not in the manual?

Keithley 2002 Specifications

General IconGeneral
BrandKeithley
Model2002
CategoryMultimeter
LanguageEnglish

Related product manuals