14. Set the measurement rate to 10 PLC by sending the following command:
RES:NPLC 10
15. Set the Model 5700A output mode to OPERATE.
16. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
17. Wait until the following message is displayed on the Model 2110:
Cali OK
18. Set the Model 5700A output to 1 kΩ.
19. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
20. Wait until the following message is displayed on the Model 2110:
Cali OK
21. Set the Model 5700A output mode to STANDBY.
22. Configure the Model 2110 for 2-wire, 10 kΩ resistance measurements by sending the following
command:
CONF:RES 1E4
23. Set the measurement rate to 10 PLC by sending the following command:
RES:NPLC 10
24. Set the Model 5700A to 0Ω and the output mode to OPERATE.
25. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
26. Wait until the following message is displayed on the Model 2110:
Cali OK
27. Set the Model 5700A output to 10kΩ.
28. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
29. Wait until the following message is displayed on the Model 2110:
Cali OK
30. Set the Model 5700A output mode to STANDBY.
31. Configure the Model 2110 for 2-wire, 100 k resistance measurements by sending the following
command:
CONF:RES 1E5
32. Set the measurement rate to 10 PLC by sending the following command:
RES:NPLC 10
33. Set the Model 5700A to 0Ω and the output mode to OPERATE.
34. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
35. Wait until the following message is displayed on the Model 2110:
Cali OK