EasyManuals Logo
Home>Keithley>Multimeter>2110

Keithley 2110 Calibration Manual

Keithley 2110
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Section
3: Performance verification Model 2110 5½ Digit Multimeter
Calibration Manual
3-4 2110-905-01 Rev. C / August 2013
AC current accuracy verification
AC CURRENT ACCURACY VERIFICATION
Range Test Point Accuracy Limits
Upper Limit Lower Limit
1 A
0.90000 A @ 60 Hz
0.90220 A
0.89780 A
0.90000 A @ 5 kHz
0.90220 A
0.89780 A
3 A
2.70000 A @ 60 Hz
2.7099 A
2.6901 A
2.70000 A @ 5 kHz
2.7099 A
2.6901 A
10 A
9.0000 A @ 60 Hz
9.0570 A
8.9430 A
9.0000 A @ 5 kHz
9.0570 A
8.9430 A
Specifications valid for 2-hour warm-up at 5-1/2 digits:
a. Slow ac filter (3 Hz bandwidth).
b. Pure sine wave input greater than 5% or range.
Frequency accuracy verification
FREQUENCY ACCURACY VERIFICATION
Range Test Point Accuracy Limits
Upper Limit Lower Limit
10-40 Hz
20.0000 Hz @ 1 Vpp
20.006 Hz
19.994 Hz
40-300 kHz
200.000 kHz @ 1 Vpp
200.04 kHz
199.96 kHz
Specifications valid for 2-hour warm-up at 5-1/2 digits:
a. Slow ac filter (3 Hz bandwidth).
b. Pure sine wave input greater than 5% of range.

Other manuals for Keithley 2110

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2110 and is the answer not in the manual?

Keithley 2110 Specifications

General IconGeneral
BrandKeithley
Model2110
CategoryMultimeter
LanguageEnglish

Related product manuals