In this section:
Verification information ............................................................. 2-1
Verification equipment .............................................................. 2-1
Verification information
In order to ensure that the power supply is working properly and performance accuracy, we
recommend that you use the equipment listed for verifying your equipment. Also, note the following
conditions and parameters that you should follow when performing verification procedures.
It is recommended that you verify instrument specifications when:
Using the power supply in a new environment.
Removing the cover for any reason.
The following are the environmental conditions that are required when performing verification and
adjustment:
Indoor location only; no direct sunlight; dust free area
Relative humidity <80%
Temperature +18 to 28 °C (64 to 82 °F)
Warm-up time 30 minutes
If the verification fails:
You will need to perform the adjustment procedure.
If the adjustment does not accomplish the desired result you will have to send the instrument to
your local Keithley Instruments office, sales partner, or distributor for repair (see
Contact
Information (on page 1-1).
Section 2
Verification preparation

Other manuals for Keithley 2260B-30-72

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2260B-30-72 and is the answer not in the manual?

Keithley 2260B-30-72 Specifications

General IconGeneral
BrandKeithley
Model2260B-30-72
CategoryPower Supply
LanguageEnglish

Related product manuals