*See full specications for other pulse regions.
Carefully consider and congure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to
a device that can deliver energy. Failure to
consider the output-off state, source levels, and
compliance levels may result in damage to the
instrument or to the device under test.
Carefully consider and congure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to
a device that can deliver energy. Failure to
consider the output-off state, source levels, and
compliance levels may result in damage to the
instrument or to the device under test.
Power and environmental specications
For indoor use only.
Power Supply 100 VAC to 240 VAC, 50 Hz to 60 Hz
(auto sensing)
Maximum VA 240 VA
Operating altitude Maximum 2000 m (6562 feet) above
sea level
Operating
temperature
0 °C to 50 °C (32 °F to 122 °F),
70% relative humidity up to 35 °C.
Derate 3% relative humidity,
35 °C to 50 °C
Storage temperature 25 °C to 65 °C (77 °F to 149 °F)
Pollution degree 1 or 2
DC source output 40.4 W maximum
±40 VDC maximum
±1.01 A at ±40 VDC
±3.03 A at ±6 VDC
Pulse region 4* Region maximums: 10 A at 20 V
Maximum pulse width: 1.8 ms
Maximum duty cycle: 1%
Measurement input Measurement category I
Voltage: 40 VDC maximum HI to LO,
250 VDC LO to ground
Current: 3.03 A at 6 V; 1.01 A at 40 V
Impedance: Variable