EasyManuals Logo
Home>Keithley>Multimeter>3706A

Keithley 3706A User Manual

Keithley 3706A
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Section
5: Typical applications Model 3762 10-Channel High Voltage Multiplexer Card
User's Manual
5-2 3762-900-01 Rev. A / July 2017
Guarded application
Non-driven guarded measurements are made with the set up shown in the next figure. The jumper is
installed to provide a guard that completely encloses HI and LO. This guards HI and LO from other
channels on the card for improved channel isolation.
Figure 9: Guarded floating measurement
Driven guard application
If multiple sources are switched to a DUT using a driven guard, as shown in the next figure, the guard
must be switched. The jumper for each channel in use is removed so that the various sources will not
try to drive the guards. Instrument LO is not switched or is switched by another card in the next two
figure.
Figure 10: 3762 Single pole switch with switched guard
Note that the previous figure labeled B is the same configuration for connecting one source to
multiple DUTs. In this case, the jumper is removed to reduce the capacitive loading on the guard.
LO and GUARD are shorted at both ends of the signal path to allow for the switched guard. LO for the
sources and the DUT must have an external connection.

Other manuals for Keithley 3706A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3706A and is the answer not in the manual?

Keithley 3706A Specifications

General IconGeneral
BrandKeithley
Model3706A
CategoryMultimeter
LanguageEnglish

Related product manuals