APPLE II (APPLE Interface)
The following program obtains one reading from the Model 617 Elac-
trometar and displays the reading on the APPLE II screan, using an
APPLE IEEE-488 interface.
DIRECTIONS
1. Using the front panel program feature, set the primary address of the
Model 617 to 27.
2. Connect the Model 617 to the APPLE II and APPLE IEEE-488 inter-
face.
3. Enter the following program using the RETURN key after each line.
4. Type in RUN and depress the RETURN key.
5. The display will read “TEST SETUP”.
6. To program the Model 617 to the &A range and take a reading, type
in FlR7TlX and depress the RETURN key.
7. The display will read NDCA + O.WOOOE-06 when the Model 617 is
in zero check.
PROGRAM
COMMENTS
20 20 = CHRW6)
30 INPUT’ITEST SETlJP?“;B$
40 PRW
50 INW
60 PRINT “RA”
70 PRINT ‘W”.ZGB$
I, ,
80 PRINT “LFl”
90 PRINT “RD”; CHRSL91);ZS;:
INPUT ” “;A$
100 PRINT “UT”
110 PRKl
120 IN#O
130 PRINT A$
140 GO TD 30
Terminator.
Enter programming command.
Example: 2& range=FlRTTlX
Send output to IEEE bus.
Get input from IEEE bus.
Turn remote on.
Write 88 to 617.
tinefeed on.
Read data from 617.
Send output to CRT.
Get input from keyboard.
Repeat.
NOTE: If conversion to numeric variable is needed, add the following:
134 A=VAL(MID$lA$,5,16))
136 PRINT A
45