2400S-903-01 Rev. E / September 2011 Return to Section Topics 1-15
Series 2400 SourceMeter® Quick Start Guide Section 1: Introduction
NOTE The five element data string is the GPIB default condition.
SCPI commands
SCPI commands to source-measure and measure only are provided in Table 6, and the
commands to measure ohms are provided in Table 7.
Programming examples
Source-measure example — Table 8 shows a typical command sequence to source 10V and
measure current on the 10mA range.
Table 1-6
SCPI commands; source-measure and measure only
1
Command
2
Description
:SOURce:FUNCtion <name> Select source function; <name> = VOLTage or CURRent.
:SOURce:xxx:MODE FIXed Select fixed V or I sourcing mode.
:SOURce:xxx:RANGe <n> Select measurement range for V or I source; <n> = range.
:SOURce:xxx:LEVel <n> Set source amplitude; <n> = amplitude in volts or amps.
[:SENSe]:FUNCtion <name> Select measure function; <name> = “VOLTage” or “CUR-
Rent”.
[:SENSe]:xxx:PROTection <n> Set V or I compliance; <n> = compliance.
[:SENSe]:xxx:RANGe <n> Select V or I measure range; <n> = range.
[:SENSe]:xxx:RANGe:AUTO
<b>
Enable or disable auto range; <b> = ON or OFF.
:OUTPut <b> Turn output on or off; <b> = ON or OFF.
:READ? Trigger and acquire one data string.
1. For measure only, the following rules apply (Table 9 provides an example to measure I only):
a. To measure V only, you must source I. To measure I only
, you must source V.
b. The source must be set to 0V or 0A.
c. When measuring V only, DO NOT use auto range.
d. When measuring V only, voltage compliance MUST be set higher th
an the volt
age being measured.
2. :xxx = :VOLTage or :CURRent.
Table 1-7
SCPI commands; measure ohms
Command Description
[:SENSe]:FUNCtion “RESistance” Select ohms function.
[:SENSe]:RESistance:MODE <name> Select ohms mode; <name> = AUTO or MANual*.
[:SENSe]:RESistance:OCOMpensated
<b>
Enable/disable offset compensation.
[:SENSe]:RESistance:RANGe <n> Select range for auto ohms; <n> = range.
[:SENSe]:RESistance:RANGe:AUTO
<b>
Enable or disable auto range for auto ohms; <b> =
ON or OFF.
:OUTPut <b
> Turn output on or off; <b> = ON or OFF.
:READ? Trigger and acquire one data string.
*For manual ohms, you must configure the SourceMeter to source-measure. Table 11 provides a programming
example to measure ohms using the manual ohms method.

Other manuals for Keithley SourceMeter 2400

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2400 and is the answer not in the manual?

Keithley SourceMeter 2400 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals