TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2011 95
14 Error messages
ERR01
Unstable weight value or impossible resetting.
Check environmental conditions.
ERR02 Adjustment error, e.g. unstable environmental conditions.
ERR03 Adjustment error, e.g. incorrect adjustment weight.
ERR04 Too low/unstable unit weight.
ERR05
Data transmission impossible as the weight value is unstable.
Check environmental conditions.
ERR06
Unstable weight value in the density determination mode.
Check environmental conditions.
ERR07
Erroneous data upload (counting the number of pieces, density
determination etc.).
ERR08 Internal adjustment error.
“UNLOAD”
Weighing range exceeded.
Check the scale plate position.
“CAL But”
Carry out scale adjustment.
Weighing range exceeded (upwards), the load placed exceeds
the scale weighing range. Remove the load from the scale.
Weighing range exceeded (downwards), e.g. the scale plate
not placed.
15 Maintenance, service and disposal
Before you start any works related to the maintenance, cleaning and
repair, disconnect the device from the operating voltage.
15.1 Cleaning
Do not use any aggressive cleaning agents (solvents etc.), but clean the device with
a cloth and mild soap solution. The liquid must not get inside the device. After the
cleaning is finished, wipe the device dry using a soft cloth.
Any loose specimen/powder remains can be removed carefully with a brush or a
handheld vacuum cleaner.
Remove any scattered weighed material immediately.
15.2 Maintenance and service
The device can be operated and maintained solely by the technicians trained
and authorized by KERN.
Disconnect from the mains before opening.