15.3 Disposal
The packaging and the device should be disposed in accordance with the national or
regional law in the location where the device is operated.
16 Help for any minor failures
If there are any programme execution problems, the scale should be switched off and
disconnected from the mains for a while. Next, the weighing process should be
started anew.
Help:
Problem Possible cause
The weight indicator is not lit
The scale is not on.
Interrupted mains connection (mains cable not
connected/damaged).
Mains voltage failure.
Batteries/rechargeable batteries
not placed correctly or discharged
No batteries/rechargeable batteries.
The weight indication
keeps fluctuating.
Draft / air movements.
Table/air vibrations.
The scale plate is in contact with foreign bodies.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the
interfering device).
The weighing result
is clearly wrong.
The scale indication was not reset.
Incorrect adjustment.
There are heavy temperature fluctuations.
The heating time not observed.
Electromagnetic fields / static discharge (select
another location / if possible, switch off the
interfering device).
If there are any other error messages displayed, switch the scale off and then on
again. If the error message persists, contact the manufacturer.

Table of Contents

Other manuals for KERN ALS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN ALS Series and is the answer not in the manual?

KERN ALS Series Specifications

General IconGeneral
BrandKERN
ModelALS Series
CategoryScales
LanguageEnglish

Related product manuals