9.3.3 AU PC
The weighing values are sent automatically and continuously, no matter if the value is
stable or unstable.
e. Format for stable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
B
*
B B B B B B B B 0 . 0 B g B B C
R
LF
f. Format in case of fault
1 2 3 4 5 6 7 8 9 10 11 12 13 14
B B B B B B B E r r o r CR LF
g. Format for unstable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
B
*
B B B B B B B B 0 . 0 B CR LF
9.3.4 rE Cr
The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.
Take into account that the following remote control commands must be sent
without a subsequent CR LF.
s Function: Stable weighing value for the weight is sent via the RS232
interface
w Function: Weighing value for the weight (stable or unstable) is sent via the
RS232 interface
t Function: No data are sent, the balance carries out the tare function.
h. Format for stable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
B
*
B B B B B B B B 0 . 0 B g B B C
R
LF
i. Format in case of fault
1 2 3 4 5 6 7 8 9 10 11 12 13 14
B B B B B B B E r r o r CR LF
j. Format for unstable values for weight/quantity/percentage
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
B
*
B B B B B B B B 0 . 0 B CR LF