EasyManuals Logo
Home>KERN>Scales>EMS 300-3

KERN EMS 300-3 User Manual

KERN EMS 300-3
31 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
30 EMS-BA-e-1716
10 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible cause
The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted.
Batteries are inserted incorrectly or empty
No batteries inserted.
The
displayed weight is permanently
changing
-
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with
foreign matter.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing value is obviously
wrong
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EMS 300-3 and is the answer not in the manual?

KERN EMS 300-3 Specifications

General IconGeneral
BrandKERN
ModelEMS 300-3
CategoryScales
LanguageEnglish

Related product manuals