TFKB_B-BA-e-2210 19
Place the adjustment weight and confirm with button,
< wait > followed by < reMvld > will be displayed.
Once < reMvld > is displayed, remove the adjustment
weight.
After successful adjustment the balance automatically
returns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong >.
Switch off balance and repeat the adjustment process.
7.9.2 External adjustment with user-defined adjustment weight < caleud >
Press and hold the TARE and ON/OFF buttons
simultaneously to enter the setup menu.
Wait until the first menu item <Cal> is displayed.
Confirm by button, < CalExt> will be displayed.
Use the navigation keys to select < caleud >.
Acknowledge by button. The numeric input window
for the weight value of the adjustment weight appears.
The active digit is flashing.
Provide adjustment weight.
Enter weight value, numeric input see chap. 3.2.2