EasyManuals Logo
Home>KERN>Scales>FKB 8K0.1

KERN FKB 8K0.1 User Manual

KERN FKB 8K0.1
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
TFKB_B-BA-e-2210 37
11.2 Using the application
11.2.1 Piece counting
Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
The higher the reference quantity the higher the counting exactness.
Especially high reference must be selected for small parts or parts with
considerably different sizes.
Smallest counting weight see table „Technical data“.
1. Set reference
Reference quantity 5, 10, 20 or 50
If necessary, put on and tare the weighing container.
Put on the desired quantity of reference pieces.
Invoke menu setting < ref > and confirm by button.
Use the navigation keys to select the reference quantity
(5, 10, 20, 50) according to the placed reference and
confirm with the button.
The balance will calculate the average item weight and
then displays the quantity of pieces.
Remove reference weight. The balance is now in piece
counting mode counting all units on the weighing plate.

Table of Contents