9
The internal light comes on when the door opens and go off when it is shut. It can be changed
so the light is either on or off all the time. To change the setting, access the menu in the usual
way by holding both the arrow icons/buttons and scroll through the menu until you get to the
Lt screen or hold the light icon/button which will take you straight to the Aut screen.
Touch the tick icon/button as before to access this function and the screen will show Aut. This is the
default setting described above. To change the setting to the light being on all the time touch an
arrow icon/button until the screen displays on. Once selected, hold the tick icon/button until the
display reverts to Lt. To change the light operation to not coming on at all, select OFF instead and
again hold the tick icon/button until the screen reverts to Lt. You can either touch the mute icon/
button or just leave the controller alone and the screen will return to the normal temperature display.
10
If you do not t batteries you will need to turn this alarm to OFF. To do so, hold the arrow
icons/buttons to enter the menu. Scroll through until you get to the screen that shows bAt
and touch the tick icon/button. Touch the arrow icon/button to change the screen from on
to OFF and hold the tick icon/button until the display reverts to bAt to conrm the change.
The battery symbol should then disappear from the temperature display screen and the alarm
won't sound.
There is a secondary temperature probe at the back of your refrigerator. It is a ‘wandering
probe’ you can move anywhere in the fridge, for example placing it in a simulated load. Like
probe t1 the controller will alarm if probe t2 detects that the temperature is lower than 2oC
or higher than 8C where it is placed. The temperature detected by this probe will display in
normal operation after the t1 screen. You can tell which one is which because a small t1 or t2
will appear above the temperature displayed.
Likewise you will be able to tell which probe is alarming as the high or low alarm will be
prexed by t1 or t2.
Internal Light
Battery Alarm
The battery alarm is enabled by default. This means if you put 4AA batteries (not supplied)
in your refrigerator the alarm will not sound unless the batteries are faulty or removed. For
extra security, the controller will alarm if the batteries are low or removed. Labcold strongly
recommends that you t batteries to your IntelliCold® pharmacy refrigerator for maximum
cold chain safety, however, should you decide not to you will need to disable the alarm, please
see Turning the Battery Alarm O opposite
The alarm will show on the normal display as a battery symbol which will alternate with a
screen showing n-b. The alarm will sound when n-b is displayed. This alarm can be muted by
touching the mute icon/button while the n-b screen is displayed. The visual alarm will remain
and the screen n-b will still alternate with the temperature display. Unless you install batteries
the alarm will sound again after the time out interval is exceeded. (see page 8).
Battery Alarm
Turning the Battery Alarm O
Temperature Probe t2
Should you wish to disable it, enter the menu by holding the two arrow icons/buttons
simultaneously and scroll until you get to a screen that shows t2.
Touch the tick icon/button to enter the t2 menu and the screen will display on. Touch the
arrow icon/button to reveal the OFF screen and hold the tick icon/button to conrm. The
display will then return to t2 and then the normal temperature display.
Temperature detected
by probe t1
Temperature detected
by probe t2