16 / 34 Puridest Glass Distillation Apparatus 09/2020
11 Technical Data
Puridest Glass Distillation Apparatus models PD 2 G, PD 4 G and PD 8 G
Exterior dimensions (W x D x H)
approx. 2.2 µS / cm at 25 °C
in conformity with DAB, bacteria and pyrogen
free, low gas content.
approx. 2.2 µS / cm at 25 °C
in conformity with DAB, bacteria and pyrogen
free, low gas content.
The conductivity of the distilled water is directly related to the chemical composition of the raw water. Components of the raw water having the
same or a lower evaporation point than water may deteriorate the conductivity.
Electromechanical temperature limiter, with ca-
pillary tube sensor.
Switch-off temperature 135 °C / -15 K
Electromechanical temperature limiter, with ca-
pillary tube sensor.
Switch-off temperature 135 °C / -15 K
Water pressure min. / max.
Mains supply
Internal fuse
On-site fuse
230 V +/-10 %, 50 / 60 Hz, 1.5 kW
Mains connection cable with Schuko plug,
internal 2 x 8 A inert, on-site 10 A inert
230 V +/-10 %, 50 / 60 Hz, 3.0 kW
Mains connection cable with Schuko plug,
internal 2 x 15 A inert, on-site 16 A inert
Caution! Mains voltage deviations, even with the mentioned limits, influence the quantity of distillate produced.
Protection class / - type
Use only inside buildings (not in potentially explo-
sive areas).
Use only inside buildings (not in potentially ex-
plosive areas).
max. 80 % rel. humidity to 31 °C, decreasing to
40 % rel. humidity at 40 °C
max. 80 % rel. humidity to 31 °C, decreasing to
40 % rel. humidity at 40 °C
Weight with water filling