EasyManuals Logo
Home>Lufft>Accessories>V200A-UMB

Lufft V200A-UMB User Manual

Default Icon
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
Operating Manual V25/09.2019
Ventus / Ventus-X / V200A
Chapter 12 Configuration and Test 26
12.2.6 Heating Parameters
Figure 12: Heating Parameters
Setpoint temperature: Temperature to which the
Ventus
is heated
default: 15°C
Heater mode: off: heating is off
auto: automated control to setpoint temperature
testmode:
Ventus
is heated to test setpoint temperature
ctrl high: heating off if control line on high,
otherwise: automatic control
ctrl low: heating off if control line on low
otherwise: automatic control
default: auto
Capacity: full: full heating power
alternating: upper plate and lower plate alternatingly
default: full
Disable if overvoltage:
the heating is disabled if overvoltage (28V) is detected
default: yes
Disable if undervoltage:
the heating is disabled if undervoltage (20V) is detected
default: no
12.2.7 Wind Parameters
Figure 13: Wind Parameters
Min. speed: starting speed of the wind sensor in m/s
default: 0,1m/s
Counts of measurements:
Number of measured values used for determining the statistical values
of average, minum value and maximum value
default: 60
Direction Offset: Correction of the wind direction
It is possible to have an offset added to the measured wind direction,
in order to convert e.g. 0° (North) into 180° (South)
default: 0°
Measurement rate: Update interval, adjustable in 250ms, 1 … 10 s
default: 10s
12.2.8 Pressure Parameters
Pressure offset: Absolute offset (for on-site adjustment) on the measured value

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lufft V200A-UMB and is the answer not in the manual?

Lufft V200A-UMB Specifications

General IconGeneral
BrandLufft
ModelV200A-UMB
CategoryAccessories
LanguageEnglish

Related product manuals