EasyManuals Logo
Home>Metrohm>Measuring Instruments>FOSS NIRS XDS MultiVial Analyzer

Metrohm FOSS NIRS XDS MultiVial Analyzer User Manual

Metrohm FOSS NIRS XDS MultiVial Analyzer
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
▪▪▪▪▪▪▪
3
Change Control
(initiated with version 1.0):
Version Date Summary of Changes
1.0
April 15,
2005
Initial Release of Manual.
1.01
August 4,
2005
Replaced photos in section 5.4 on Iris Adapter, replaced photos in
section 5.5, revised text to clarify correct software settings for use
with cells.
1.02 June 8, 2009
Revised format of Table of Contents
Moved Wavelength Linearization before Reference Standardization.
Replaced section on XDS Instrument Connection with an updated,
more current version. This also includes basic connection
troubleshooting. The sections on Windows 95, 98, and NT 4.0 were
dropped.
Added note that USP dropped “Photometric Linearity Test” as of
12-2008.
Changed photo in Lamp Replacement, now showing arrow on
lamp.
Corrected reference to Lamp Replacement in Troubleshooting to
section 8.3.
Added note to Troubleshooting, Instrument Calibration, that
Reference Standardization must be performed.
Removed discussion of sample spacers in section 5.4. Spacers are
not necessary with this unit. The slight optical difference in height
between “Use Iris” mode and “Use Tray” mode is handled by
having a separate Reference Standardization for each mode.
Changes were made in Reference Standardization to clarify this
issue. It was previously described; however additional information is
now given. Removed mention of spacers from page 9, and from
Index.
Added note about wavelength drift due to temperature change
with the WSR standard. This is on the page immediately following
Instrument Calibration.
1.03
June 12,
2009
Inserted specific instructions for Reference Standardization in “Use
Iris” position.
1.04
June 25,
2009
Added notes that Reference Standardization must be repeated after
changing spot size.
Added warning not to run with cover open, which can cause X-Y
Positioning Errors.
Added information about “Shipping Position” and foam used to
protect mechanism.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm FOSS NIRS XDS MultiVial Analyzer and is the answer not in the manual?

Metrohm FOSS NIRS XDS MultiVial Analyzer Specifications

General IconGeneral
BrandMetrohm
ModelFOSS NIRS XDS MultiVial Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals