EasyManuals Logo
Home>Metrohm>Measuring Instruments>FOSS NIRS XDS MultiVial Analyzer

Metrohm FOSS NIRS XDS MultiVial Analyzer User Manual

Metrohm FOSS NIRS XDS MultiVial Analyzer
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
70
▪▪▪▪▪▪▪
Vision requests the “Standard File” for the Certified
Wavelength Standard.
Place the mini-CD into the CD drive, and select the
standard file. Click “Open”.
Vision prompts the user to insert the WSR10xxx
standard into the instrument.
Insert the WSR10Xxx Wavelength Standard Cell when
prompted. Use the tray provided for this purpose.
Insert the standard with the label parallel to the long
axis of the tray.
Click “OK” to continue.
This test takes about 45 seconds.
At the end of the test this dialog box is displayed.
Print confirmation of test results if needed for a log
book or other file.
Click “OK” to exit Instrument Calibration.
Hint: This file may be manually copied on the
computer, in the Vision directory, if desired.
During Instrument Calibration, the wavelength response for each defined peak is adjusted, to assure
precise wavelength registration between instruments. At the same time, bandwidth (bandpass) is
measured, and is iteratively adjusted to an optimum value for the peaks measured. This is performed
to assure good agreement from instrument to instrument, should multiple instruments be used for
analysis of similar products.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm FOSS NIRS XDS MultiVial Analyzer and is the answer not in the manual?

Metrohm FOSS NIRS XDS MultiVial Analyzer Specifications

General IconGeneral
BrandMetrohm
ModelFOSS NIRS XDS MultiVial Analyzer
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals