2 - 3
(1) Pad expiration window: checks the expiration date of pads.
(2) Latch: opens or closes the lid.
(3) Handle
(4) Status indicator
• Green: the equipment is turned on, and can work correctly.
• Flashing green: the equipment is in the standby status, and is ready for
operation at any time.
• Flashing red: auto test failure is detected on the equipment.
• Off: no battery is installed or the battery is malfunctioning.
(5) Speaker: the equipment automatically adjusts the volume depending on
surrounding noise levels by default.
(6) Display screen (for equipment configured with the screen)
(7) Pads connector: connects the electrode pads.
(8) Pads package holder: stores the electrode pads.
(9) Adult/Child mode switch: flip right or left to switch between adult and child.
(10) Language button: press to switch between the configured languages.
(11) Optical sensor (for equipment configured with the screen):
the equipment automatically adjusts the screen brightness depending on
surrounding light by default.
(12) Shock button (for semi-auto model): press to deliver a shock to the patient.
(13) Microphone: records voices. It is available only when the record function is
enabled.