A - 5
A.9 Wireless Specifications
A.10 Defibrillator Specifications
Wi-Fi
Standard IEEE 802.11 a/b/g/n
Operating frequency IEEE 802.11 b/g/n (at 2.4G): 2.412 GHz to 2.472 GHz
IEEE 802.11 a/n (at 5G): 5.18 GHz to 5.24 GHz, 5.745 GHz to 5.825 GHz
Data security Standards: WPA-PSK, WPA2-PSK, WPA-Enterprise,WPA2-Enterprise
EAP method: EAP-TLS, PEAP-GTC, PEAP- MSCHAPv2
Encryption: TKIP, AES
Modulation mode DSSS and OFDM
Cellular
Operating frequency LTE-FDD B1: 1920 MHz to 1980 MHz, 2110 MHz to 2170 MHz
LTE-FDD B3: 1710 MHz to 1785 MHz, 1805 MHz to 1880 MHz
LTE-FDD B7: 2500 MHz to 2570 MHz, 2620 MHz to 2690 MHz
LTE-FDD B8: 880 MHz to 915 MHz, 925 MHz to 960 MHz
LTE-FDD B20: 832 MHz to 862 MHz, 791 MHz to 821 MHz
LTE-FDD B28A:703 MHz to 733 MHz, 758 MHz to 788 MHz
LTE-TDD B38: 2570 MHz to 2620 MHz
LTE-TDD B40: 2300 MHz to 2400 MHz
Standard/Modulation mode 3GPP E-UTRA Release 11: LTE-FDD/LTE-TDD
Standards Meet standards of IEC 60601-2-4
Defibrillation mode BeneHeart C1/BeneHeart C1A/BeneHeart C2/BeneHeart C2A/
BeneHeart S1/BeneHeart S1A/BeneHeart S2/BeneHeart S2A:
semi automatic external defibrillation
BeneHeart C1 Fully Automatic/BeneHeart C1A Fully Automatic/
BeneHeart C2 Fully Automatic/BeneHeart C2A Fully Automatic/
BeneHeart S1 Fully Automatic/BeneHeart S1A Fully Automatic/
BeneHeart S2 Fully Automatic/BeneHeart S2A Fully Automatic/:
fully automatic external defibrillation
Defibrillation waveform Biphasic truncated exponential (BTE) waveform, auto-compensation
according to patient impedance
Defibrillation electrodes Multifunction electrode pads.
Range of selected energy For adults: 100 J, 150 J, 170 J, 200 J, 300 J, 360 J.
For children: 10 J, 15 J, 20 J, 30 J, 50 J, 70 J, 100 J.
Patient impedance range 25 to 300 Ω
Shock series Energy level: 100 to 360J, configurable for adults.
10 to 100J, configurable for children.
Shocks: 1, 2, 3, configurable;
Meeting AHA/ECR guidelines 2015 by default.
ECG Analysis Performance See BMindray Shockable Rhythm Analysis Algorithm.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart C2 Fully Automatic and is the answer not in the manual?

Mindray BeneHeart C2 Fully Automatic Specifications

General IconGeneral
BrandMindray
ModelBeneHeart C2 Fully Automatic
CategoryMedical Equipment
LanguageEnglish