8.3.2 Auto Test
The equipment with a battery installed carries out auto test at the configured time even when powered off to
check the equipments operational performance and alert the operator if a problem exists. Auto test is initiated
at 3:00 am every day by default.
The equipment provides no voice prompts during auto test. The test result can be checked on the status
indicator:
Flashes in green: the auto test passes, an auto-test report is saved automatically when the test is
completed.
Flashes in red: the auto test fails. If the equipment is connected to the AED ALERT system, an auto-test
report is saved and uploaded automatically to the system when the test is completed.
Mindray recommends you check the status indicator every day, and record the result according to GInspection
Record.
CAUTION
With the equipment powered off, auto test can be performed only when the battery is installed.
NOTE
When the equipment is placed at a temperature below -20°C, the auto test cannot be carried out,
and incorrect status indication could result.
8.3.3 Electrode Pads Check
The expiration date of electrode pads should be checked every month. You can check the expiration date from
the pad expiration window, and record it according to GInspection Record.
8.4 Disposing the Equipment
Dispose of the equipment and its accessories when its service life is reached. Follow local regulations regarding
the disposal of such product.
WARNING
For disposal of parts and accessories, where not otherwise specified, follow local regulations
regarding disposal of hospital waste.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart C2 Fully Automatic and is the answer not in the manual?

Mindray BeneHeart C2 Fully Automatic Specifications

General IconGeneral
BrandMindray
ModelBeneHeart C2 Fully Automatic
CategoryMedical Equipment
LanguageEnglish

Related product manuals