3 - 3
3.3 Turning on the Equipment
Before turning on the equipment, perform the following inspections:
Check for mechanical damage on the equipment or other damage on the pads package.
Make sure the pads cable is properly connected and battery installed.
Check the expiration date of the pads on the pads package.
Open the AED lid, then the equipment automatically powers on.
3.4 Switching Voice Language
You can press the Language button until the desired language is selected. At most three voice languages can be
configured.
3.5 Turning off the Equipment
Before turning off the equipment, perform the following inspections:
1. Confirm that the patient therapy is completed.
2. Disconnect the electrode pads from the patient.
To turn off the equipment, close the AED lid.
WARNING
If the patient is not connected to the equipment, and no operation is found performed on the
equipment within 30 minutes, the equipment will automatically shut down.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart S1 Fully Automatic and is the answer not in the manual?

Mindray BeneHeart S1 Fully Automatic Specifications

General IconGeneral
BrandMindray
ModelBeneHeart S1 Fully Automatic
CategoryMedical Equipment
LanguageEnglish

Related product manuals