EasyManuals Logo
Home>Mindray>Medical Equipment>cPM 12

Mindray cPM 12 User Manual

Mindray cPM 12
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
6-9
6.6.12 Module Failures
Symptoms Possible Cause Corrective Action
Failed to connect the
external parameter modules
Module defective 1. Check that the cable between the external converter
board inside the module and the converter board is
correctly connected,
2. Replace the converter board.
Main unit defective 1. Check that the cable between the main board and
power management board is correctly connected.
2. Replace the power management board.
3. Replace the main board.
Module can be loaded, but
"XX communication
stopped" is reported or some
parameters cannot be used
Cable defective inside the
module
Check the cables connecting the converter board and
corresponding parameter module.
Parameter module defective Replace the corresponding module.
Converter board defective
inside the module
Replace corresponding converter board.
6.6.13 Technical Alarm Messages
Refer to the Passport 12/Passport 8 Patient Monitor Operator's Manual (PN: 046-004417-00) and cPM 12/cPM 8
Patient Monitor Operator's Manual (PN: 046-011732-00).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray cPM 12 and is the answer not in the manual?

Mindray cPM 12 Specifications

General IconGeneral
BrandMindray
ModelcPM 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals